HEAD OFFICE 07 38651139  CHINCHILLA 1800 234 477

Additel 780S Pressure Controller

Additel ADT780S Pressure ControllerThe Additel ADT780S controller incorporates the same durability, accuracy, and quality into a new bench top controller packed with features and functionality that is remarkably easy to use.

The ADT780S offers two base ranges: to 1,000 psi (70 bar) and to 3,000 psi (200 bar). The base range establishes the maximum controlling range of the controller. Each configuration includes a control sensor in which you can preselect the sensor range best suited your application. External and internal sensors can be used which allows for expanded range and accuracy capability in the future. 

The 780S Pressure Controller is considered to handle a very wide range of applications which may normally require a pressure calibrator. 

Features of the ADT780S

  • Pressure ranges from  vacuum to 3,000 psi (200 bar)
  • Removable interchangeable intelligent sensors
  • Precision accuracy of 0.005% of reading plus 0.005% of full scale 
  • Standalone solution to 3,000 psi (200 bar), no gas bottle required when used with the Electric Pump and 3K Gas Booster
  • Standalone solution to 1,000 psi (70 bar), no gas bottle required when used with the Electric Pump
  • Standard model accuracy of 0.02% of full scale
  • External pressure modules to 3,000 psi (200 bar)
  • WiFi enabled communications
  • Fully temperature compensated accuracy over 0°C to 50°C
  • Large 7” color touch screen display
  • Control stability of 0.003%FS
  • Built-in barometer
  • Easy-to-use icon based user interface

 

Measurement Specifications

Model Number ADT780S-1K ADT780S-3K
Pressure range -14.5 to 1,000 psi   (-0.95 to 70 bar) -14.5 to 3,000 psi   (-0.95 to 200 bar)(1)
Control stability 0.003% FS
Precision (includes 1 year stability) See pressure range table
1 Year stability 0.01% FS standard accuracy sensors 0.005% FS precision accuracy sensors
Media clean  Gas
Over-range indication 103% to 120% (based on sensor)
Resolution 4, 5, 6, or 7 digits (user selectable)(2)
Pressure type Gauge, Absolute
Warm up time 15 minutes

Typical pressure settling time (within 0.003%FS)

20 seconds (10%FS pressure change   assuming 50ml test volume)
Measurement units Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C,cmHg@0°C,   mHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@4°C, cmH2O@4°C, mH2O@4°C,  mmH2O@20°C, cmH2O@20°C, mH2O@20°C, kg/m2,   kg/cm2, mtorr, torr, atm, lb/ft2, tsi, user selectable
Minimum control pressure(3) 0.0001 psi 0.001 psi
 
(1). Gas booster or HP gas supply required to reach 3,000 psi (200 bar). Booster ordered separatel.
(2). Digit resolution for precision model only.
(3). Dependent on pressure module.

 

Barometric Measurement Specifications

Sensor/Mode Gauge Absolute
ADT160A-02-CPXX N/A 40Pa(1)
ADT160A-01-APXXQ 3Pa(2) N/A

 

Barometer range (60~110)KPa, the accuracy is 40Pa (11Pa is optional)

(1). 40Pa uncertainty (k=2) includes calibration uncertainty, linearity and long term stability (< 30Pa per year). Barometer range of 60 to 120KPa.

(2). Combined linearity, hysteresis and repeatability. Add 3Pa when used in gauge mode. When using the ADT160A-01-APXXQ sensors with the ADT780 controller in gauge mode, regular zeroing will realign the barometric reading to provide the most accurate result.

  

Physical Specifications

Specification ADT780S-1K & ADT780S-3K
Power 100 – 240V,  50/60 Hz
Pressure ports G1/8 F
Storage temperature -20°C to 70°C
Operating environment 0-95% RH non-condensing
Display 7 inch (17.8 cm) color, touch screen   display
Weight 33 lbs (15 kg)
Dimensions(DWH) 16.5 x 17.3 x 5.2 inch (419 x 440 x   132 mm)
Mounting Standard desktop, optional rack mount
Shock 4G
Vibration 1G   10Hz~500Hz

 

Electric Pump and Booster

ADT780S-1K-EPUMP Electric Pump

Generate Vacuum to 1,000 psi(700bar) 
Built-in filter and liquid trap

Pressure range -13.5 to 1,080 psi (-0.95 to 75 bar)
Weight

69.6 lbs (31.6   kg)

Size(WHD) 17.3 X 9.3 X 21.7 inch   (440 X 235X 550 mm)
Power 100 to 240V
Media Air
Outlet port G   1/8F
Storage temperature -20°C to 70°C
Operating environment 0-90% RH   non-condensing
Mounting Standard   desktop, optional rack mount
Control operation 2   LED displays with pressure limit settings
Typical max pressure time Approx   2 min with ADT780 Controller

  

ADT780S-3K-BST Booster 
3,000 psi (200bar) Gas booster specification

Minimum inlet pressure 300 psi (20 bar)
Out pressure range 300 to 3,600 psi (250 bar)
Inlet flow rate 500ml/min
Size (WHD) 17.8 X 7.3 X 23.9 inch   (451 X 185 X 607 mm)
Weight 74 lbs (33.6   kg)
Power 100 to 240V
Supply media Nitrogen or clean gas

 

The Range of ADT160A Intelligent Digital Pressure Modules

P/N Pressure Range Media Accuracy Pressure Rating Suggested Controller Compatibility Internal/ External Use
psia bar.a Burst Over Pressure

Precision Quartz Sensors[1]

AP45Q 0 to 45 3.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both
AP100Q 0 to 100 7.0 G 0.005% rdg + 0.005% FS 3x 1.2x ADT780-1K only Both
AP400Q 0 to 400 28 G 0.005% rdg + 0.005% FS 3x 1.2x Both Both
AP500Q 0 to 500 35 G 0.005% rdg + 0.005% FS 3x 1.03x Both Both
AP1KQ 0 to 1,000 70 G 0.005% rdg + 0.005% FS 3x 1.1x Both Both
AP2KQ 0 to 2,000 140 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both
AP3KQ 0 to 3,000 200 G 0.005% rdg + 0.005% FS 3x 1.1x ADT780-3K only Both

Standard Pressure Sensors

P/N Pressure Range Media Accuracy Pressure Rating Suggested Controller Compatibility Internal/ External Use
psig bar.g Burst Over Pressure
CP10 ±10 ±0.7 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP15 ±15 ±1.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP30 -15 to 30 -1 to 2.0 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP50 -15 to 50 -1 to 3.5 G 0.02% FS 3x 1.2x ADT780-1K only Both
CP100 -15 to 100 -1 to 7.0 G,L 0.02% FS 3x 1.2x ADT780-1K only Both
CP300 -15 to 300 -1 to 20 G,L 0.02% FS 3x 1.2x Both Both
CP500 -15 to 500 -1 to 35 G,L 0.02% FS 3x 1.2x Both Both
CP600 -15 to 600 -1 to 40 G,L 0.02% FS 3x 1.2x Both Both
CP1K -15 to 1,000 -1 to 70 G,L 0.02% FS 3x 1.2x Both Both
CP2K -15 to 2000 -1 to 140 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
CP3K -15 to 3,000 -1 to 200 G,L 0.02% FS 3x 1.2x ADT780-3K only Both
CP5K -15 to 5000 -1 to 350 G,L 0.02% FS 3x 1.2x N/A External
CP10K -15 to 10,000 -1 to 700 G,L 0.02% FS 2x 1.2x N/A External

 

[1] Contact Additel for other range options.
**Low pressure sensors (ADT160A-XX-DPX) available for low pressure and differential pressure measurement. Also available for low pressure control (gauge mode only). Ranges from ±1 inH20 (2.5 mbar) to 300 inH2O (700 mbar).